Cook Incorporated recalls Williams Cystoscopic Injection Needles
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- What's wrong
- The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
- Products
-
- Williams Cystoscopic Injection Needles
- Company
- Cook Incorporated
- Where it was sold
-
US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.
States: Florida, Illinois, New Mexico, Texas, Washington
- Barcodes
- 00827002161122
- Units affected
- 97 units
- Status
- Ongoing
Recall Relay is not affiliated with or endorsed by FDA or any agency. Details here are copied from the FDA notice and may lag behind it; the official notice is always the final word.