Howmedica Osteonics Corp. recalls TS Misc. Instruments Tray
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- What's wrong
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- Products
-
- TS Misc. Instruments Tray
- Company
- Howmedica Osteonics Corp.
- Where it was sold
-
These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
- Barcodes
- 07613327555127
- Units affected
- 577 trays
- Status
- Ongoing