Beckman Coulter, Inc. recalls ROTOR ASSY-SW40
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- What's wrong
- Optima XE XPN ultracentrifuge rotors may contain a microstructure defect that can cause rotors to break into two or more pieces (bifurcate) during a centrifugation run, resulting in destruction of samples being processed. When used for in vitro diagnostic (IVD) sample preparation, rotor breakage may delay patient test results. Additionally, sample destruction may expose laboratory personnel to biohazardous materials.
- Products
-
- ROTOR ASSY-SW40
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
US Nationwide distribution.
- Units affected
- 92 units
- Status
- Ongoing