Beckman Coulter, Inc. recalls ROTOR ASSY-SW41

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Read the official notice on accessdata.fda.gov
What's wrong
Rotors have been found with a non-conforming microstructure, which may result in some rotors breaking into two or more pieces during use. Rotors that break into multiple pieces during use may result in significant damage to materials (i.e., samples) being processed in the centrifuge and to the centrifuge instrument.
Products
  • ROTOR ASSY-SW41
Company
Beckman Coulter, Inc.
Where it was sold

US Nationwide distribution.

Units affected
336 units