BioMerieux SA recalls EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrume
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- What's wrong
- Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
- Products
-
- EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrume
- Company
- BioMerieux SA
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
States: California, Indiana, Kansas, Louisiana, Minnesota, Missouri, New York, Ohio, Texas
- Units affected
- 33
- Status
- Ongoing
Recall Relay is not affiliated with or endorsed by FDA or any agency. Details here are copied from the FDA notice and may lag behind it; the official notice is always the final word.