BioMerieux SA recalls EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrume

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Read the official notice on accessdata.fda.gov
What's wrong
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
Products
  • EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrume
Company
BioMerieux SA
Where it was sold

Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.

States: California, Indiana, Kansas, Louisiana, Minnesota, Missouri, New York, Ohio, Texas

Units affected
33
Status
Ongoing