Boston Scientific Neuromodulation Corporation recalls Nalu TD2 Therapy Disc with firmware

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Read the official notice on accessdata.fda.gov
What's wrong
Therapy Disc (TD) that powers and sends stimulation signals to an Implantable Pulse Generator (IPG), part of a neurostimulation system, has a firmware issue that causes the TD to attempt to restart, then fail, leading it to become non-functional, and if a secondary TD is unavailable, this may lead IPG to become non-functional, thus unable to inhibit transmission of pain signals.
Products
  • Nalu TD2 Therapy Disc with firmware
Company
Boston Scientific Neuromodulation Corporation
Where it was sold

US Nationwide distribution in the states of PA, FL, OR, AZ, OH, SC, TX, OR, OK, MI, CA, NJ.

States: Arizona, California, Florida, Michigan, New Jersey, Ohio, Oklahoma, Oregon, Pennsylvania, South Carolina, Texas

Units affected
134
Status
Ongoing