Stryker Leibinger GmbH & Co. KG recalls Precision Pointer Electromagnetic (8000-050-001)
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- What's wrong
- Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
- Products
-
- Precision Pointer Electromagnetic (8000-050-001)
- Company
- Stryker Leibinger GmbH & Co. KG
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.
States: Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington
- Barcodes
- 07613327413205
- Units affected
- 513
- Status
- Ongoing