Auris Health Inc recalls
Auris Health Inc is named on 4 recalls in Recall Relay's records since November 2023, from FDA (7 recalled products: FDA lists each product as its own recall); the most recent was January 21, 2026.
All 7 recalled products, newest first.
-
Class IIofficial
Auris Health, Inc recalls MONARCH PlatformDevice for bronchoscopic visualization, patient airway access has software issue: if application restarts after patient-side selection, prior to bronchoscope loading, application will re-initialize to patient-left position; and if right position was previously selected, this may lead to unanticipated robotic arm positioning toward patient-left, may result in contact with patient/endotracheal tube.
-
Class IIofficial
Auris Health, Inc recalls MONARCH Bronchoscope. Model Number: MBR-000211-BPotential that product was leak tested with equipment outside of its expected operating range, resulting in bronchoscopes that may not be susceptible to leaks.
-
Class IIofficial
Auris Health Inc recalls Monarch Platform (Urology)Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
-
Class IIofficial
Auris Health Inc recalls Monarch Platform (Endoscopy)Endoscopy/bronchoscopy/urology devices may have a strain relief defect with some cart and tower power cords that may cause an electrical short, which may lead to system shutdown, which can lead to prolonged procedure or case conversion, and if the user (e.g., operating room staff) touches exposed wires or conductive fluids in contact with the exposed wires, there is a chance of electrical shock.
-
Class IIofficial
Auris Health Inc recalls Monarch Platform REF MON-000005-01Their is the potential that software issues may result in flipped image of the virtual Bronchoscope view.
-
Class IIofficial
Auris Health Inc recalls Monarch Platform REF MON-000005-01RTheir is the potential that software issues may result in flipped image of the virtual Bronchoscope view.
-
Class IIofficial
Auris Health Inc recalls Monarch Platform REF MON-000006 100-240V- 50/60Hz 1440wTheir is the potential that software issues may result in flipped image of the virtual Bronchoscope view.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.