Foundation Medicine, Inc. recalls
Foundation Medicine, Inc. is named on 7 recalls in Recall Relay's records since November 2023, from FDA (8 recalled products: FDA lists each product as its own recall); the most recent was June 10, 2026.
All 8 recalled products, newest first.
-
Class IIofficial
Foundation Medicine, Inc. recalls Hamilton Microlab STAR Liquid Handler; Protocol Configuration Tool (PCT) v2.0.0. Used to conduct F1CDx and F1LSoftware issue that can cause an increase in spontaneous software closures.
-
Class IIofficial
Foundation Medicine, Inc. recalls Protocol Configuration Tool (PCT) software v1.3.0 for Hamilton AutoLys STAR used within FoundationOneCDx (F1CDPotential for spontaneous closures of Hamilton HxRunControl software during execution of Autolys protocols after deployment of the PCT software for AutoLys Hamilton STAR.
-
Class IIofficial
Foundation Medicine, Inc. recalls Sequencing Agent SEQ0067Possible increase in phasing, potentially resulting in false-positive BARD1 splice site indel artifact variants
-
Class IIofficial
Foundation Medicine, Inc. recalls FoundationOne Companion Diagnostic (F1CDx)Four (4) total reports were sent to customers with clinically relevant variants requiring amendment, one (1) of which with an incorrect negative CDx claim. The remaining three (3) with tumor profiling result changes did not impact CDx indications.
-
Class IIofficial
Foundation Medicine, Inc. recalls FoundationOne Companion Diagnostic (F1CDx)Two (2) reports were sent to customers with an incorrect negative claim on the claims page; the device variant information was displayed correctly in the tumor profiling section of both FDA-approved test reports.
-
Class IIofficial
Foundation Medicine, Inc. recalls FoundationOne Companion Diagnostic (F1CDx)An incorrect negative claim was identified on the claims page; the device variant information was displayed correctly in the tumor profiling section of the FDA-approved test report.
-
Class IIofficial
Foundation Medicine, Inc. recalls FoundationOne Liquid CDx (F1LCDx)Reports were sent to customers without the companion diagnostic (CDx) Claims Page. Claims pages were distributed via amended reports within 7 days of the issue occurrence.
-
Class IIofficial
Foundation Medicine, Inc. recalls FoundationOne CDx (F1CDx)Reports were sent to customers without the companion diagnostic (CDx) Claims Page. Claims pages were distributed via amended reports within 7 days of the issue occurrence.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.