Globus Medical, Inc. recalls
Globus Medical, Inc. is named on 4 recalls in Recall Relay's records since November 2023, from FDA (6 recalled products: FDA lists each product as its own recall); the most recent was April 3, 2025.
All 6 recalled products, newest first.
-
Class IIofficial
Globus Medical, Inc. recalls MAGEC 2 RodDevices for the affected lot were not assembled according to product specifications. The distraction rod outer diameter was assembled with a 5.5mm rod versus the required 5.0mm.
-
Class IIofficial
Globus Medical, Inc. recalls ExcelsiusGPS Cannula ADevices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.
-
Class IIofficial
Globus Medical, Inc. recalls Globus Medical Cranial Drill with StopSterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
-
Class IIofficial
Globus Medical, Inc. recalls Globus Medical Cranial Drill with StopSterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
-
Class IIofficial
Globus Medical, Inc. recalls Globus Medical Cranial DrillSterile-packed Cranial Drills may not be sterile due to a breach in the packaging and/or may have discoloration near the drill tip.
-
Class Iofficial
Globus Medical, Inc. recalls ExcelsiusGPS Flat Panel Fluoroscopy Fixture KitA calibration error associated with the Flat Panel Fluoroscopy Fixture when used with ExcelsiusGPS for fluoroscopy workflow may lead to inaccurate navigation, which may result in device misplacement.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Globus Medical, Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.