Heraeus Medical GmbH (Dental Division) recalls
Heraeus Medical GmbH (Dental Division) is named on 3 recalls in Recall Relay's records since November 2023, from FDA (10 recalled products: FDA lists each product as its own recall); the most recent was August 20, 2026.
All 10 recalled products, newest first.
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls PALACOS R pro 40; PALACOS R pro 80. Model/Catalog Numbers: 5081286; 5081287. RadiopaquePotential for monomer liquid leakage.
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls PALACOS R+G pro 40; PALACOS R+G pro 80. Model/Catalog Number: 5081288; 5081289. RadiopaquePotential for monomer liquid leakage.
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls PALACOS MV pro 40; PALACOS MV pro 80. Model/Catalog Number: 5150633; 5150635. RadiopaquePotential for monomer liquid leakage.
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls PALACOS MV+G pro 40; PALACOS MV+G pro 80. Model/Catalog Number: 5081290; 5081291. RadiopaquePotential for monomer liquid leakage.
-
Class IIIofficial
Heraeus Medical GmbH (Dental Division) recalls HeraeusBased on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
-
Class IIIofficial
Heraeus Medical GmbH (Dental Division) recalls HeraeusBased on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Soincrease of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS MV pro 40; PALACOS MV pro 80 Model/Catalog Number: 5150633; 5150635increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
-
Class IIofficial
Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS R+G pro 40; PALACOS R+G pro 80 Model/Catalog Number: 5081288; 508128increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Heraeus Medical GmbH (Dental Division) or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts