Johnson & Johnson Surgical Vision, Inc. recalls
Johnson & Johnson Surgical Vision, Inc. is named on 3 recalls in Recall Relay's records since November 2023, from FDA; the most recent was July 31, 2026.
All 3, newest first.
-
Class IIofficial
Johnson & Johnson Surgical Vision, Inc. recalls HEALON EndoCoat PROSurgical instructions sent for viscoelastic products used during ophthalmic viscosurgical cataract procedures including in part: If resistance is encountered during insertion, withdraw cannula and reassess the insertion conditions (i.e. insertion angle, incision size, etc.) before proceeding because forcing the cannula may cause corneal abrasion, iris trauma or descemet's membrane detachment.
-
Class IIofficial
Johnson & Johnson Surgical Vision, Inc. recalls Johnson & Johnson VISION Reusable LAMINAR High Flow Irrigation Sleeve and Test Chamber 21 Gauge REF OPOHF21L TDue to a manufacturing variation of the irrigation sleeves which could result in a missing port hole, this non-conformity could lead to insufficient flow to cool the ultrasonic phaco tip, potentially leading to an unstable anterior chamber which could cause possible harms.
-
Class IIofficial
Johnson & Johnson Surgical Vision, Inc. recalls TECNIS Toric II OptiBlue IOL Models ZCWDue to intraocular lens containing an angle out of specification (OOS). The cylinder axis marks do not meet product specifications for the allowable angle between Low Power Meridian (LPM) and toric cylinder axis marks.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Johnson & Johnson Surgical Vision, Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.
Get recall alerts