Megadyne Medical Products, Inc. recalls
Megadyne Medical Products, Inc. is named on 3 recalls in Recall Relay's records since November 2023, from FDA (8 recalled products: FDA lists each product as its own recall); the most recent was June 17, 2024.
All 8 recalled products, newest first.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls Brand Name: MEGADYNE" MEGA SOFT" Dual Reusable Patient Return Electrode Product Name: MEGADYNE" MEGA SOFT" DuaTo limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls Brand Name: MEGADYNE" MEGA 2000" Patient Return Electrode Product Name: MEGADYNE" MEGA 2000" Patient Return ElTo limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls Brand Name: MEGADYNE" MEGA SOFT" Reusable Patient Return Electrode Product Name: MEGADYNE" MEGA SOFT" ReusableTo limit the use of the Mega Soft Pads to patients aged 12 or older to mitigate the potential risk of patient burns identified with use of Mega Soft pads.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls MEGADYNE MEGA SOFT Pediatric Patient Return Electrode. Reusable Patient Return Electrode for patients weighingReports of patient burns.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls Product Code 0845Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls Product Code 0846Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls Product Code 0847Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
-
Class Iofficial
Megadyne Medical Products, Inc. recalls Product Code 0848Affected product codes are now limited to use in patients age 12 years or older. Megadyne has received reports of patient burn injuries up to and including third-degree burns requiring intervention which may lead to prolonged hospital stay, scarring, and additional surgeries in both pediatric and adult patients.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.