NxStage Medical Inc recalls
NxStage Medical Inc is named on 2 recalls in Recall Relay's records since November 2023, from FDA; the most recent was April 28, 2025.
All 2, newest first.
-
Class IIofficial
NxStage Medical Inc recalls NxStage PureFlow-B Solution. Catalog numbers: RFP-400Potential for the perimeter seal on the small chamber of the two-chamber bag to fail causing the electrolyte fluid to splash into the face of the user when attempting to open the peel seam to mix the two chambers prior to treatment.
-
Class IIofficial
NxStage Medical Inc recalls NxStage PureFlow-B SolutionThe smaller chamber of the two chamber bag is bursting when nurses are attempting to mix the two chambers prior to use. The electrolyte fluid in the small chamber of the NxStage PureFlow B Solution has a pH of less than 2.0. The solution can be harmful if it comes in contact with the eyes or skin, and/or as a result of a slip or fall injury due to spilled fluid.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions NxStage Medical Inc or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.