Physio-Control, Inc. recalls
Physio-Control, Inc. is named on 10 recalls in Recall Relay's records since November 2023, from FDA (21 recalled products: FDA lists each product as its own recall); the most recent was April 13, 2026.
All 21 recalled products, newest first.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Some defibrillators serviced between July 2023 through November 2025 were serviced without objective evidence that a Performance Inspection Procedure (PIP) was conducted, and the inability to perform the device's intended use may result in potential harms such as delays or inability to provide therapy.
-
Class IIofficial
Physio-Control, Inc. recalls PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI code:Due to pediatric defibrillator electrode delamination
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 1000 Part / Catalog Numbers / UDI-DI code: 320371500023 / 99425-000025 / 885074698953 320371500229 / 9Due to required inspections not being performed on products/units that have gone through servicing.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15 Part / Catalog Numbers / UDI-DI code: V15-7-000022 / 99577-001957 / 883873871690 V15-2-000051 / 995Due to required inspections not being performed on products/units that have gone through servicing.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK CR2 Part / Catalog Numbers / UDI-DI code: CR2-2-001451 99512-001434 883873785911 LIFEPAK CR2: The LIFEDue to required inspections not being performed on products/units that have gone through servicing.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 20e REF 99507-000001 The AED mode is to be used only on patients in cardiopulmonary arrest.Incorrect keypad was incorrectly installed onto a defibrillator/monitor.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 35 ECG cable REF 11111-000041Product shipped with incorrect IFU. IFU does not include proper cleaning/disinfection instructions
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo¿ RD rainbow Adt 8¿ SpCODue to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15 (LP15) Model/Catalog Numbers: o 97577-000101 o 99577-000046 o 99577-000055 o 99577-000061 o 99577-0Due to an out of tolerance tool being used on monitor/defibrillator systems.
-
Class IIofficial
Physio-Control, Inc. recalls STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP20 STERILIZABLE INTERNAL DEFIBRILLATION PADDLES REF LP15 DeDue to customer complaints and out of box failures, the firm is conducting a voluntary recall as a result of product failures (cracks and over-molding separation of the sterilizable internal defibrillation paddles.
-
Class IIofficial
Physio-Control, Inc. recalls AC Power Adapter for use with the LIFEPAK 15 Monitor/DefibrillatorAC Power Adapter (ACPA) may not charge the monitor/defibrillator batteries and may not allow the monitor/defibrillator to power on while plugged into the ACPA.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
-
Class IIofficial
Physio-Control, Inc. recalls LIFEPAK 15Their is a potential for battery pins to become loose within the defibrillator resulting in power issues.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.