Roche Diagnostics Operations, Inc. recalls
Roche Diagnostics Operations, Inc. is named on 5 recalls in Recall Relay's records since November 2023, from FDA; the most recent was January 20, 2026.
All 5, newest first.
-
Class IIofficial
Roche Diagnostics Operations, Inc. recalls cobas pro integrated solutions with cobas c 503 analytical units: cobas pro sample supply unitSoftware defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
-
Class IIofficial
Roche Diagnostics Operations, Inc. recalls The Elecsys Anti-TSHR immunoassay is a three-step competitive immunoassay with streptavidin-coated microparticIssues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
-
Class IIofficial
Roche Diagnostics Operations, Inc. recalls cobas HCYSAbnormal calibration signals and control recovery issues on the Homocysteine Enzymatic (HCYS) Assay can produce erroneous patient results (positive or negative bias)
-
Class IIofficial
Roche Diagnostics Operations, Inc. recalls Creatine Kinase (CK) used on cobas c 311/501/502 and COBAS INTEGRA 400 plus-In vitro test for the quantitativeCreatine Kinase (CK) Assay Abnormal calibrations on the cobas¿ c 311/501/502 and COBAS INTEGRA 400 plus Analyzers may lead to erroneous results and misdiagnosis.
-
Class IIofficial
Roche Diagnostics Operations, Inc. recalls Tina-quant D-Dimer Test SystemElevated results were detected with Li-heparin plasma samples when compared to citrated plasma samples.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.