Securitas Healthcare LLC recalls
Securitas Healthcare LLC is named on 6 recalls in Recall Relay's records since November 2023, from FDA (9 recalled products: FDA lists each product as its own recall); the most recent was December 8, 2025.
All 9 recalled products, newest first.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial 900 MHz Call StationSecuritas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial 900 MHz Call StationSecuritas Healthcare has become aware of a firmware issue in the Call Stations in the Affected Lots that may cause Call Stations in the Affected Lots to (1) not transmit a low battery alert to the Arial Wireless Emergency Call System prior to battery failure, or (2) not transmit a low battery alert to the Arial Wireless Emergency Call system at least seven (7) days prior to the battery failing.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial Server Software Versions 11.1.4 through 11.1.6Certain software versions will not notify end users of calls initiated by call points with the Arial Emergency and Nurse Call System.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial Call StationThis is a risk that alarms will not be received by the Arial Wireless Nurse Call System and premature depletion of the battery.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial Call StationThere is a risk that alarms will not be received by the Arial Wireless Emergency Call System and premature depletion of the battery.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial mobile applicationThe Arial Mobile Application utilizing Apple iOS 17.4 and iOS 17.5 is frequently restarting causing the app to receive alerts for all pending alarm notifications and other events when the app restarts, experience difficulty in placing and maintaining staff-to-staff phone calls, and frequently appears offline while the app restarts.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial Water Resistant Pendant with Neck LanyardPremature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial Water Resistant Pendant with WristbandPremature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.
-
Class IIofficial
Securitas Healthcare LLC recalls Arial Water Resistant Pendant with Long WristbandPremature battery depletion presents a risk that alarms will not be received by the Arial Wireless Emergency Call System.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.