United Orthopedic Corp. recalls
United Orthopedic Corp. is named on 3 recalls in Recall Relay's records since November 2023, from FDA (9 recalled products: FDA lists each product as its own recall); the most recent was May 20, 2026.
All 9 recalled products, newest first.
-
Class IIofficial
United Orthopedic Corp. recalls Brand Name: U2 Total Knee System Product Name: Femoral valgus offset adaptorDue to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
-
Class IIofficial
United Orthopedic Corp. recalls Brand Name: USTAR II Knee System Product Name: Hinge Box TrialDue to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
-
Class IIofficial
United Orthopedic Corp. recalls Brand Name: U2 Total Knee System Product Name: PSA Box TrialDue to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
-
Class IIofficial
United Orthopedic Corp. recalls Brand Name: U2 Total Knee System Product Name: Femoral valgus adaptorDue to an incomplete process of the component screw by a subcontractor which prevents the properly assembling of stem trail.
-
Class IIofficial
United Orthopedic Corp. recalls Brand Name: U2 Total Knee System Product Name: Femoral offset bushingDue to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.
-
Class IIofficial
United Orthopedic Corp. recalls Brand Name: U2 Total Knee System Product Name: Femoral offset bushingDue to a design issue causing the incorrect engraved numbering on implant that could result in the incorrect orientation of the implant.
-
Class IIofficial
United Orthopedic Corporation recalls USTAR II Knee System Cemented curved stemTheir is a potential that the implant curved stem may be oriented incorrectly.
-
Class IIofficial
United Orthopedic Corporation recalls USTAR II Hip System Press-Fit Curved StemTheir is a potential that the implant curved stem may be oriented incorrectly.
-
Class IIofficial
United Orthopedic Corporation recalls USTAR II Knee System : Cemented Curved StemTheir is a potential that the implant curved stem may be oriented incorrectly.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.