Vortex Surgical Inc. recalls
Vortex Surgical Inc. is named on 3 recalls in Recall Relay's records since November 2023, from FDA (15 recalled products: FDA lists each product as its own recall); the most recent was December 16, 2025.
All 15 recalled products, newest first.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical Lindsell Sutured IOL MarkerThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Rumex Disposable Diamond Dusted ILM ElevatorThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Volk Single Use Vitrectomy Lenses 1. Flat Vitrectomy LensThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Disposable Forceps and Cannula 1. 23GA ACTU8 Forceps Adaptive VS0740.23 2. 25GA ACTU8 Forceps AdaptiveThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls 25GA Subretinal Injection Cannula VS0220.25There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Tecfen Retractable Membrane PolisherThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical 25GA I.D.D. Internal Delivery DeviceThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Oertli 1. 23G Laser Probe OS4 Flex-Tip VK401113 2. 23G Laser Probe OS4 Illuminated Flex-Tip VK401213 3. 23G LaThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical 25GA BackflushThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical I2 Injection KitThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical TID Pharos Illuminated DepressorXXX
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical 1. 23GA Laser Probe Curved Cat No. VS0120.23 2. 25GA Laser Probe Curved Cat No. VS0120.25 3. 2There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.25GThere may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical 25ga Illuminated Flex-Tip Laser ProbeReason for the voluntary recall is some products in this lot are not passing through a 25ga cannula smoothly.
-
Class IIofficial
Vortex Surgical Inc. recalls Vortex Surgical ACTU8 FORCEPSReason for the voluntary recall is residue was identified on the two forceps tips of a customer in Korea. The residue is associated with one lot of forceps tips which were utilized in two lots of ACTU8 Adaptive Forceps Tips occurring in 0.25% of those lots.
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.
Severity labels say where they come from: “official” is the agency's own class, “advisory” is based on the agency's safety warning, and “est. risk” is Recall Relay's estimate from the hazard description, not an agency classification.
Hear about the next one
Recall Relay can email you when a new recall mentions Vortex Surgical Inc. or any words you choose. Start with a free key (no card needed), then set up alerts on your account page.