ZOLL Medical Corporation recalls
ZOLL Medical Corporation is named on 3 recalls in Recall Relay's records since November 2023, from FDA (10 recalled products: FDA lists each product as its own recall); the most recent was February 13, 2025.
All 10 recalled products, newest first.
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80A; Software Version: NA Product DThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80CThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-80AThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-00AThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-00A-TSOThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5S-01CThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: (1) G5A-01CThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
ZOLL Medical Corporation recalls Brand Name: ZOLL Product Name: Powerheart G5 AED Model/Catalog Number: G5S-80C; Software Version: NA Product DThe potential for devices to fail their self-test as a result of prolonged exposure to extreme environments (temperature and/or humidity).
-
Class IIofficial
Zoll Medical Corporation recalls ZOLL Powerheart G5 AEDG5 Semi-Automatic AED is shipped, with a protective film over its front panel, to protect the screen and shock button from cosmetic damage during shipping. It has come to our attention that customers may not be removing the protective film during deployment of the product. If the protective film is left adhered to the front bezel it may prevent the user from actuating the shock button and may lead to a delay or prevent delivery of defibrillation therapy to a victim suffering from sudden cardiac arrest
-
Class Iofficial
ZOLL Medical Corporation recalls ZOLL 731 Ventilator (EMV+Operator's Guide & Quick Reference Guide (QRG) ZOLL 731 Ventilator for MRI Compatible Devices. Patient Safety Ventilator MRI information omitted from the manuals. Update to the instructions ensures safe distance from the MRI to ensure proper function of the ventilator and reduce the risk of delay in ventilator therapy
This page groups recalls by the company name the agency printed, so it can include different companies with similar names, and a recall may name a store or distributor rather than the maker. It counts the recalls in Recall Relay's records, not a company's whole history; an FDA recall that lists several products counts once. Open a recall to see its details and the agency's official notice, which is always the final word. Recall Relay isn't affiliated with any agency or with the companies named here.