CPSC, Fluke Corp. Announce Recall of Digital Multimeters
Estimated risk
CPSC doesn't classify recalls by risk, so Recall Relay estimated the risk as high from the hazard description. This is our estimate, not an official classification.
- Hazard
- The recalled units can take longer than normal, up to 18 seconds, to display readings of AC voltages above 500 volts. Users can misinterpret the delayed reading to mean that high voltage is not present. If high voltage is present, users could be exposed to a risk of shock, electrocution, and thermal burns.
- What to do
- Consumers should stop using these testers immediately and contact Fluke for information on returning the recalled unit for a free repair.
- Products
-
- Fluke Digital Multimeters
- Company
- Fluke Corp.
- Where it was sold
-
Home and hardware stores and industrial distributors nationwide sold these testers, from January 2001 through October 2001, for between $170 and $215.
Nationwide
- Units affected
- About 40,000
- About this recall
- Multimeters are used to measure voltage, resistance and current. The recalled Fluke digital multimeters measure up to 1,000 volts alternating current (VAC) and direct current (VDC). "Fluke" and "175", "177" or "179" are written on the front of the unit. The tester body is yellow and black, and measures about 7.25 inches long by 3.5 inches wide. Recalled units have a serial number below 79000000. The serial number is written on the back of the instrument, under the hinged stand.