Ossur H / F recalls Miami J Select Collar

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.
Products
  • Miami J Select Collar
Where it was sold

US: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK, PR

States: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Montana, North Carolina, Nebraska, New Hampshire, New Jersey, New Mexico, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington

Units affected
21,020
About this recall
Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101