Ossur H / F recalls Miami J Select Collar
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Pressure sensitive adhesive hooks attached to the sides of the anterior cervical orthosis collar panel can occasionally exhibit poor adhesion, which may lead to them gradually peeling off with repeated doffing, possibly leading to reduced immobilization of the cervical spine.
- Products
-
- Miami J Select Collar
- Company
- Ossur H / F
- Where it was sold
-
US: NH, MA, NY, NJ, PA, GA, MD, NC, IN, FL, AR, TN, KY, OH, MI, IL, TX, MO, CO, UT, AZ, WA, OR, CA, NM, VA, IA, MN, NE, SC, MT, LA, ID, SD, AL, CT, OK, AK, PR
States: Alaska, Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Iowa, Idaho, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Maryland, Michigan, Minnesota, Missouri, Montana, North Carolina, Nebraska, New Hampshire, New Jersey, New Mexico, New York, Ohio, Oklahoma, Oregon, Pennsylvania, Puerto Rico, South Carolina, South Dakota, Tennessee, Texas, Utah, Virginia, Washington
- Units affected
- 21,020
- About this recall
- Miami J Select Collar, Catalog: MJS-101; Miami J Select Collar Set, Catalog: MJSR-101