Philips North America recalls IntelliVue G7m Anesthesia Gas Module
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- The 866173 lntelliVue G7m Anesthesia Gas Modules could have a component issue that may trigger an interruption of the gas measurement. This issue concerns the G7m's sample gas pump which could stop working, producing a technical, INOP (inoperable) alarm. The pump may fail prematurely when there is friction caused by an abrasion. This recall is for an IMPORTANT PRODUCT NOTICE dated 1/18/23.
- Products
-
- IntelliVue G7m Anesthesia Gas Module
- Company
- Philips North America
- Where it was sold
-
US Nationwide. Global Distribution.
Nationwide
- Barcodes
00884838051492- Units affected
- 8,449 devices
- About this recall
- IntelliVue G7m Anesthesia Gas Module, Product Number 866173