Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Smiths Medical ASD Inc. recalls CADD Medication Cassette Reservoir

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Certain CADD Medication Cassette Reservoirs may exhibit a weakened weld joint between the medication bag and tubing due to a production equipment malfunction. This could result in a medication leakage.
Products
  • CADD Medication Cassette Reservoir
Where it was sold

Worldwide distribution.

Nationwide

Units affected
125928 units
About this recall
CADD Medication Cassette Reservoir, 100mL, Product Code 21-7002-24