Getinge Disinfection Ab recalls Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
- Products
-
- Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
- Company
- Getinge Disinfection Ab
- Where it was sold
-
Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.
States: Colorado, Massachusetts, New Hampshire, New Mexico
- Barcodes
07340153700109- Units affected
- 902 units
- About this recall
- Getinge 88-Series Washer- Disinfector. Model Number: 88-5.