Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Getinge Disinfection Ab recalls Getinge 88-Series Washer- Disinfector. Model Number: 88-5.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Potential for device to overheat during operation if the circulation pressure calibration was not performed or had been performed incorrectly, and the low-pressure alarm is not triggered.
Products
  • Getinge 88-Series Washer- Disinfector. Model Number: 88-5.
Where it was sold

Worldwide - US Nationwide distribution in the states of Colorado, Massachusetts, New Mexico, New Hampshire and the countries of Algeria, Australia, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Austria, Hong Kong, Hungary, India, Iraq, Ireland, Israel, Italy, Japan, Kenya, Korea, Republic of, Kuwait, Lithuania, Mexico, Netherlands, Norway, New Zealand, Poland, Qatar, Romania, Saudi Arabia, Singapore, Spain, Sweden, Thailand, Turkey, United Arab Emirates, United Kingdom.

States: Colorado, Massachusetts, New Hampshire, New Mexico

Barcodes
07340153700109
Units affected
902 units
About this recall
Getinge 88-Series Washer- Disinfector. Model Number: 88-5.