Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America Llc recalls IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 a

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Devices with Option CP2 and initially programmed to software version P.01.01 were configured incorrectly at the factory. Therefore, Enhanced ECG Capabilities provided by Option CP2 were not enabled in these devices. Without Option C01 Full Arrhythmia, the device will not provide the yellow alarms for enhanced arrhythmia detection. This is a retrospectively reported recall from 6/14/23.
Products
  • IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 a
Where it was sold

Foreign Only: Denmark France Germany Japan United Kingdom

Barcodes
00884838038769
Units affected
153 units OUS
About this recall
IntelliVue Patient Monitor MX400/MX450/MX500/MX550 with software version P.01.01 and ordered with Option CP2 are affected as follows: 1) IntelliVue MX400 866060; 2) IntelliVue MX450 866062; 3) IntelliVue MX500 866064; 4) IntelliVue MX550 866066 Indicated for use by health care professionals whenever there is a need for monitoring the physiological parameters of patients.