Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Philips North America recalls Patient Information Center iX

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.
Products
  • Patient Information Center iX
Where it was sold

Worldwide - US Nationwide distribution.

Nationwide

Barcodes
00884838104594
Units affected
591 devices
About this recall
Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1