Philips North America recalls Patient Information Center iX
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- During extended operation of the MX40 with the PIC iX, the DHCP (Dynamic Host Configuration Protocol) lease can expire on the MX40. If the DHCP lease expires while the device is offline, upon reconnecting with the PIC iX, a new IP address will be assigned to the MX40. The PIC iX will not recognize the new IP address after reconnecting and therefore the expecting settings sync fails.
- Products
-
- Patient Information Center iX
- Company
- Philips North America
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
00884838104594- Units affected
- 591 devices
- About this recall
- Patient Information Center iX, Product Code 866386, and Patient Information Center iX Expand, Product Code 866390, Software Revision 4.0.1