ImaCor Inc. recalls ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
- Products
-
- ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile
- Company
- ImaCor Inc.
- Where it was sold
-
US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.
States: Florida, Georgia, New Jersey, Pennsylvania, South Carolina
- Barcodes
00861589000108- Units affected
- 45 units
- About this recall
- ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.