Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

ImaCor Inc. recalls ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.
Products
  • ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile
Where it was sold

US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.

States: Florida, Georgia, New Jersey, Pennsylvania, South Carolina

Barcodes
00861589000108
Units affected
45 units
About this recall
ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.