Baxter Healthcare Corporation recalls Welch Allyn Life2000 Ventilation System

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.
Products
  • Welch Allyn Life2000 Ventilation System
Where it was sold

US nationwide distribution.

Nationwide

Barcodes
00887761978089
Units affected
10,957 units
About this recall