Baxter Healthcare Corporation recalls Welch Allyn Life2000 Ventilation System
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- The Life2000 ventilator may fail to initiate the Low Gas Pressure alarm if the pressure gas source (Life2000 compressor, oxygen cylinder or wall source) is not supplied to the ventilator before initiating therapy.
- Products
-
- Welch Allyn Life2000 Ventilation System
- Where it was sold
-
US nationwide distribution.
Nationwide
- Barcodes
00887761978089- Units affected
- 10,957 units
- About this recall
- Welch Allyn Life2000 Ventilation System, REF MS-01-0118, within exterior carton codes: 1) BT-20-0002, 2) BT-20-0002A, 3) BT-20-0002AP, 4) BT-20-0007, 5) BT-20-0011, and 6) RMS010118CP