FDAMedical devicesRecalled Status: Ongoing

ASP Global, LLC. recalls Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 30" 2 LINE (10) Model/Catalog Number: DT-7030-02Q Soft

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to incorrect expiration date on product manufactured between September 2016 to March 2023. Products are labeled with a 10 year shelf life. The correct validated shelf life is 5 years.
Products
  • Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 30" 2 LINE (10) Model/Catalog Number: DT-7030-02Q Soft
Where it was sold

US Nationwide distribution in the states of MO, NV, IN, MN, GA, WI, PA, TX, CT, LA, NJ, FL, AL, NY, SC, NE, NC, OH, AZ, CO, MA, IL, KY, CA, VA, ME, KS, MD, TN, MS, AR, WA, NM, MI, WY.

States: Alabama, Arkansas, Arizona, California, Colorado, Connecticut, Florida, Georgia, Illinois, Indiana, Kansas, Kentucky, Louisiana, Massachusetts, Maryland, Maine, Michigan, Minnesota, Missouri, Mississippi, North Carolina, Nebraska, New Jersey, New Mexico, Nevada, New York, Ohio, Pennsylvania, South Carolina, Tennessee, Texas, Virginia, Washington, Wisconsin, Wyoming

Units affected
820
About this recall
Brand Name: MPR Product Name:SINGLE USE TOURNIQUET CUFF 30" 2 LINE (10) Model/Catalog Number: DT-7030-02Q Software Version: Not Applicable Product Description: Sterile Disposable Tourniquet Cuff is used to be wrapped around a patient's limb and inflated to reduce or totally occlude circulation during surgery. It is non-invasive device. Tourniquet Cuffs are intended for use in surgical procedures that requires the interruption of blood flow and clear visualization of the operative field. Component: No