Cardinal Health 200, LLC recalls Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Kits contain Microstream components. Medtronic issued an update to the Instructions for Use of these components due reports of difficulty in disconnecting the adapter from the patient's endotracheal tube to perform a procedure. The issue can result in a delay to treatment and unintended extubation. The potential exists for respiratory failure, hypoxia, aspiration, or low oxygen saturation.
- Products
-
- Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366
- Company
- Cardinal Health 200, LLC
- Where it was sold
-
Pending
- Barcodes
101971069500741019895617798710197106159828101971067040351019895616225910197106052105- Units affected
- 210 total kits
- About this recall
- Presource Packs containing Medtronic Microstream Components: Intubation Kit Catalog Numbers AN23366084 AN23366085 AN2336608H AN2336608J Intubating Bag Kit Catalog Number OMSVM4166A Intubation UH Kit Catalog Number AN2336608G