Beckman Coulter, Inc. recalls Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls

Class IIIofficial

FDA Class III

FDA classified this recall as Class III: the product is unlikely to cause health problems but breaks the agency's rules (a labeling mistake, for example).

Hazard
Beckman Coulter has become aware of an increase in customer complaints for intermittent IRISpec CB negative glucose control failure, resulting in false positive control results when used in conjunction with iChemVELOCITY Urine Chemistry Strips (PNs 800-7204 and 800-7212) on the iChemVELOCITY Analyzer. This issue may delay the reporting of patient results, with a worst-case scenario of a delay in diagnosis or treatment of metabolic disorders, kidney function abnormalities, urinary tract infections, and liver function.
Products
  • Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls
Where it was sold

worldwide

Nationwide

Barcodes
10837461002611
Units affected
5589 units
About this recall
Beckman Coulter IRISPEC CA/CB/CC Urine Chemistry Controls, Catalog Number REF 800-7211 and 800-7702, in vitro diagnostic device.