Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Beckman Coulter, Inc. recalls The Access TSH (3rd IS) assay is a paramagnetic particle

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Certain lots of Access TSH reagent packs may produce erroneously low TSH (thyroid-stimulating hormone) results.
Products
  • The Access TSH (3rd IS) assay is a paramagnetic particle
Where it was sold

Worldwide distribution - US Nationwide and the countries of Austria, Belgium, Croatia, France, Germany, Ireland, Italy, Slovakia, Spain, Switzerland, United Kingdom of Great Britain and Northern Ireland.

Nationwide

Barcodes
15099590744373
Units affected
2805 units
About this recall
The Access TSH (3rd IS) assay is a paramagnetic particle, chemiluminescent immunoassay for the quantitative etermdination of human thyroid-stimulating hormone (thyrotropin, TSH, hTSH) levels in human serum and plasma using the Access Immunoassay Systems. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders. This assay is capable of providing 3rd generation TSH results. The DxI 9000 Access Immunoassay Analyzer is an in vitro diagnostic device used for the quantitative, semi-quantitative, or qualitative determination of various analyte concentrations found in human body fluids.