Ventec Life Systems, Inc. recalls VC+Pro Model Number: PRT-00900-001
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
- Products
-
- VC+Pro Model Number: PRT-00900-001
- Company
- Ventec Life Systems, Inc.
- Where it was sold
-
Distribution US nationwide including Puerto Rico.
Nationwide
- Barcodes
00855573007785- Units affected
- 1,534 units
- About this recall
- VC+Pro Model Number: PRT-00900-001, Package PN:PRT-00973-000. used for ventilatory support.