Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Change Healthcare Canada Company recalls Change Healthcare Cardiology Hemo. Versions 14.2

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A potential issue has been identified in Change Healthcare Cardiology Hemo where incorrect data entry can lead to inaccurate hemodynamic calculations, potentially resulting in misdiagnosis and inappropriate treatment. This issue occurs when the unit of measurement (UOM) for hemoglobin is configured in a way that differs from the users understanding.
Products
  • Change Healthcare Cardiology Hemo. Versions 14.2
Where it was sold

Worldwide distribution - US Nationwide and the countries of Australia, Canada, Germany, Ireland, Israel, United Kingdom.

Nationwide

Barcodes
17540262100013
Units affected
221 systems
About this recall
Change Healthcare Cardiology Hemo. Versions 14.2, 14.1.1, 14.2, 14.3, 14.3.2, 14.3, 15.0.1, 15.1.