FDAMedical devicesRecalled Status: Ongoing

B-K Medical A/S recalls bk3000

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The internal power connections at the power inlet of the ultrasound devices can come loose resulting in the potential for excessive heat which could lead to a thermal event or the device becoming inoperable.
Products
  • bk3000
Where it was sold

Worldwide - US Nationwide - There was also government/military distribution and the countries of Australia, Austria, Belgium, Brazil, Bulgaria, Canada, China, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, Finland, France, Germany, Greece, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Republic of Korea, Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Morocco, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Saudi Arabia, Serbia, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Thailand, Turkey, Ukraine, United Arab Emirates, and United Kingdom.

Nationwide

Barcodes
05704916000264
Units affected
1648 devices
About this recall
bk3000, model #2300-11; bk3500, model #2300-41; bk5000, model #2300-61; and bkActive 2300, model #2300-66, ultrasound system with battery option.