Datex--ohmeda, Inc. recalls GE Healthcare Giraffe Incubator Carestation CS1
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Prior to current established manufacturing practices, higher levels of airborne formaldehyde exposure from newly manufactured incubators could be seen during clinical use that would rapidly decrease over a week. Affected customers were notified in June 2023 to run incubators for a week to potentially off gas any residual airborne formaldehyde. The general scientific consensus is that sensory irritation of the eyes is the most sensitive endpoint with respect to exposure to airborne formaldehyde and is the adverse effect clinically experienced at the lowest concentration.
- Products
-
- GE Healthcare Giraffe Incubator Carestation CS1
- Company
- Datex--ohmeda, Inc.
- Where it was sold
-
worldwide
Nationwide
- Barcodes
00840682116855- Units affected
- 224 units
- About this recall
- GE Healthcare Giraffe Incubator Carestation CS1, Infant Incubator. Model Numbers: 1) 2082844-002-01040417; 2) 2082844-002-01044956; 3) 2082844-002-01058424; 4) 2082844-002-01069017; 5) 2082844-002-01089666; 6) 2082844-002-01090474; 7) 2082844-002-01091269; 8) 2082844-002-01093489; 9) 2082844-002-01113262; 10) 2082844-002-01113711; 11) 2082844-002-01123085; 12) 2082844-002-01123918; 13) 2082844-002-01128966; 14) 2082844-002-01133042; 15) 2082844-002-01146447; 16) 2082844-002-100205; 17) 2082844-002-100749; 18) 2082844-002-100752; 19) 2082844-002-100995; 20) 2082844-002-101425; 21) 2082844-002-523548; 22) 2082844-002-529928; 23) 2082844-002-594800; 24) 2082844-002-643544; 25) 2082844-002-925841; 26) 2082844-002-965540; 27) 2082844-002-982916; 28) 2082844-002-983420; 29) 2082844-002-993905; 30) 2082844-002-995018; 31) 2082844-002-995799; 32) 2082844-002-998943; Radiant infant warmer