Baxter Healthcare Corporation recalls Novum IQ Syringe infusion system
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- Baxter is issuing an Urgent Medical Device Correction for the Novum IQ Syringe Pump. Baxter identified that after multiple downstream occlusion alarms, the pump may display an Infusion Complete alarm even though uninfused fluid remains in the syringe.
- Products
-
- Novum IQ Syringe infusion system
- Where it was sold
-
US Nationwide Distribution to states of: AR, AZ, CA, FL, GA, HI, IA, IN, KS, LA, MD, MI, MN, MO, NC, NE, NH, NJ, NY, PA, TX, UT, VA, VT, WA, WI AND WV.
Nationwide
- Units affected
- 2023 units
- About this recall
- Novum IQ Syringe infusion system, Product Code 40800BAXUS