Microgenics Corporation recalls MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Vial-to-vial variability in MAS¿ Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
Products
  • MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control
Where it was sold

Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX, CA, LA, SC, OH, MO, OR and the countries of KOR, BGD, THA, ZAF, DEU, VNM, BRA, NZL.

States: California, Louisiana, Michigan, Missouri, New York, Ohio, Oklahoma, Oregon, South Carolina, Texas, Utah

Units affected
1366
About this recall
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP