Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Cook Incorporated recalls Williams Cystoscopic Injection Needles

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
The female Luer lock adapter component in the affected device lots may not have been correctly seated due to misalignment and inadequate torquing during manufacturing. Potential events that may occur as a result of this issue include increased procedural time and inadequate treatment.
Products
  • Williams Cystoscopic Injection Needles
Where it was sold

US States: FL, IL, NM, TX, WA. Canada, France, Germany, Italy, Spain, Switzerland.

States: Florida, Illinois, New Mexico, Texas, Washington

Barcodes
00827002161122
Units affected
97 units
About this recall
Williams Cystoscopic Injection Needles, used for cystoscopic injection into the urethra, bladder neck, and bladder wall, Reference Part Number 090001-S26, Order Number (GPN) G16112