Philips North America recalls Cardiac Workstation 5000; Model Number: 860439;
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.
- Products
-
- Cardiac Workstation 5000; Model Number: 860439;
- Company
- Philips North America
- Where it was sold
-
US Nationwide distribution in the states of CA, IA, TN, TX, VA.
States: California, Iowa, Tennessee, Texas, Virginia
- Barcodes
00884838094826- Units affected
- 28 units
- About this recall
- Cardiac Workstation 5000; Model Number: 860439;