Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Medline Industries, LP recalls Medline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
Products
  • Medline surgical convenience kits labeled as: 1) NEURO-SPINE BASIN PACK
Where it was sold

US Nationwide distribution.

Nationwide

Barcodes
1019845909342510193489678222
Units affected
1900 kits
About this recall