Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Smiths Medical ASD, Inc. recalls medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Due to a manufacturing defect, users may experience inaccurate pressure monitoring, an interruption in pressure monitoring, or an inability to zero the device due to either a leak within the TranStar Disposable Pressure Transducer or from a defect within the TranStar Disposable Pressure Transducer chip.
Products
  • medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA
Where it was sold

Worldwide distribution.

Nationwide

Barcodes
50351688518123
Units affected
3080 units
About this recall
medex TranStar SINGLE 84IN SET WITH ANTI-SHUNT STOPCOCK 10/EA, Product Code REF MXG300243R1