Medtronic MiniMed, Inc. recalls CareLink Clinic

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
Products
  • CareLink Clinic
Where it was sold

US Nationwide distribution in the state of KS.

States: Kansas

Units affected
1
About this recall
CareLink Clinic, REF: MMT-7350