Techno-path Manufacturing Ltd. recalls Techno-Path Multichem AE Unassayed Level 2 REF: AE622X

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
A routine stability studies have identified a stability drift affecting the Ammonia (NH4+) analyte in specific lots of Multichem AE quality control material. Additional testing confirmed that the labelled shelf-life performance claim for the Ammonia analyte can no longer be supported for the affected lots. This issue is limited to the Ammonia analyte only. The impacted device may cause delay of issuing results to the physician. Delayed recognition and diagnosis of elevated levels may impact in timely treatment of life-threatening levels described within medical toxicity intervention decision points.
Products
  • Techno-Path Multichem AE Unassayed Level 2 REF: AE622X
Where it was sold

US Nationwide distribution in the state of Massachusetts.

States: Massachusetts

Barcodes
05391523442817
Units affected
123 units