Wipro GE Healthcare Private Ltd. recalls GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Carescape Central Station (CSCS), software version V3.0.5, can cause the system to enter a continuous reboot cycle. If this occurs, a loss of central monitoring of connected patients could result. This issue occurs only when the number of central stations connected to the network is 118, 119 or 120.
- Products
-
- GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5
- Where it was sold
-
Worldwide distribution.
Nationwide
- Barcodes
001952783433830019527834351200195278516367- Units affected
- 2365 units
- About this recall
- GE Healthcare Carescape Central Station (CSCS) V3 MAS800 Desktop with software version V3.0.5, Part Number 5867474-03; Medical Display, cathode ray tube