Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.
- Products
-
- DxI 9000 Access Immunoassay Analyzer
- Company
- Beckman Coulter, Inc.
- Where it was sold
-
Worldwide - US Nationwide distribution.
Nationwide
- Barcodes
15099590732103- Units affected
- 571 units
- About this recall
- DxI 9000 Access Immunoassay Analyzer, Part Number C11137