Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

UIH Technologies LLC recalls Shanghai United Imaging Healthcare Co.

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Products
  • Shanghai United Imaging Healthcare Co.
Where it was sold

US: OUS:

Barcodes
06971576832033
Units affected
3 units
About this recall
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 780 MD, Rx Only