Unomedical A/S recalls VariSoft infusion set
FDA Class I
FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.
- Hazard
- The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
- Products
-
- VariSoft infusion set
- Company
- Unomedical A/S
- Where it was sold
-
US Nationwide distribution in the state of CA.
States: California
- Barcodes
0570524401879205705244018785- Units affected
- 3362 units
- About this recall
- VariSoft infusion set, Model Number 1002830