Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Unomedical A/S recalls VariSoft infusion set

Class Iofficial

FDA Class I

FDA classified this recall as Class I: the most serious kind, meaning there is a reasonable chance the product could cause serious health problems or death.

Hazard
The Varisoft infusion set is used in conjunction with tandem insulin pumps. The defect reported was damage to the connector piece during manufacturing. This damage results in the connector becoming detached more easily from the insulin set than expected, requiring less force to disconnect than intended. Disconnection, when not detected, for example during sleep, leads to missed basal dosing. which in turn can result in elevated blood glucose and ketone level (nocturnal hyperglycemia). Diabetic ketoacidosis.
Products
  • VariSoft infusion set
Where it was sold

US Nationwide distribution in the state of CA.

States: California

Barcodes
0570524401879205705244018785
Units affected
3362 units
About this recall
VariSoft infusion set, Model Number 1002830