Ortho-Clinical Diagnostics, INc. recalls VITROS Chemistry Products Lipase Slides
FDA Class II
FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.
- Hazard
- Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
- Products
-
- VITROS Chemistry Products Lipase Slides
- Where it was sold
-
US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea
States: California, Colorado, Florida, Georgia, Illinois, Indiana, Kansas, North Carolina, New York, Tennessee, Texas, Vermont
- Barcodes
1075875000951010758750004683- Units affected
- 342 units
- About this recall
- VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)