Skip to content

All recalls FDA

FDAMedical devicesRecalled Status: Ongoing

Ortho-Clinical Diagnostics, INc. recalls VITROS Chemistry Products Lipase Slides

Class IIofficial

FDA Class II

FDA classified this recall as Class II: the product may cause health problems that are temporary or can be treated; the chance of serious harm is remote.

Hazard
Products 1668409 and 8297749 may contain VITROS Lipase Slides with an abnormality which could cause -46% to +86% bias in lipase results depending on how the slides are used. Falsely low levels could result in a missed or delayed diagnosis of acute pancreatitis and may lead to diagnostic confusion in the setting of acute abdominal symptoms. Falsely elevated levels would not result in patient harm but may cause diagnostic confusion and a delay in accurate diagnosis of a patient with abdominal pain.
Products
  • VITROS Chemistry Products Lipase Slides
Where it was sold

US States: CA, CO, FL, GA, IL, IN, KS, NC, NY, TN, TX, VT. Canada, France, India, Japan, South Korea

States: California, Colorado, Florida, Georgia, Illinois, Indiana, Kansas, North Carolina, New York, Tennessee, Texas, Vermont

Barcodes
1075875000951010758750004683
Units affected
342 units
About this recall
VITROS Chemistry Products Lipase Slides, Product Codes 1668409 (contains 60 Slides per Cartridge) and 8297749 (contains 18 slides per cartridge)